Faraday Balance.

FEATURES
Local Contact
GRENOBLE

Gabriel SEYFARTH

Description
GRENOBLE
Field range
GRENOBLE 35T
Temperature range
GRENOBLE 20K -> 1.3K (extension planned)
Sample size
GRENOBLE max: 1.5mm width and length, less for thickness to avoid inhomogeneous field within sample (500µm), single crystals (in case of anisotropy of the crystal, measurements are only possible with field along the easy axis, otherwise the torque will change the capacitance as well as the longitudinal movement, difficult to disentangle both contributions)
Sensitivity
GRENOBLE 5*10-7 emu (improvements ongoing), capacitive measurement
Typical experiment
GRENOBLE M(H), anomalies or quantum oscillations, absolute values of M only with subsequent calibration measurements
Sample Holder
GRENOBLE CuBe sample platform, single crystals attached by apiezon grease prototypes available, further improvements under construction
Sample environment
GRENOBLE exchange gas (He)
Examples
GRENOBLE
GRENOBLE
local contact: G. SEYFARTH
FEATURES GRENOBLE
Local Contact

Gabriel SEYFARTH

Description
Field range 35T
Temperature range 20K -> 1.3K (extension planned)
Sample size max: 1.5mm width and length, less for thickness to avoid inhomogeneous field within sample (500µm), single crystals (in case of anisotropy of the crystal, measurements are only possible with field along the easy axis, otherwise the torque will change the capacitance as well as the longitudinal movement, difficult to disentangle both contributions)
Sensitivity 5*10-7 emu (improvements ongoing), capacitive measurement
Typical experiment M(H), anomalies or quantum oscillations, absolute values of M only with subsequent calibration measurements
Sample Holder CuBe sample platform, single crystals attached by apiezon grease prototypes available, further improvements under construction
Sample environment exchange gas (He)